HP 1660C Series Service Manual page 83

Logic analyzers
Hide thumbs Also See for 1660C Series:
Table of Contents

Advertisement

Connect the HP 1662C/CS/CP or HP 1663C/CS/CP Logic Analyzer to the Pulse Generator
Testing
Connect to
Combination
HP 8133A
Channel 2 Output
1
Pod 1, channel 3
Pod 2, channel 3
Pod 3, channel 3
Pod 4, channel 3
Activate the data channels that are connected according to one of the previous
3
tables.
a Press the Format key.
b Select the field showing the channel assignments for one of the pods being tested.
Press the Clear entry key. Using the arrow keys, move the selector to the data
channels to be tested, then press the Select key. An asterisk means that a channel is
turned on. When all the correct channels of the pod are turned on, press the Done
key. Follow this step for the remaining pods.
To test the single-clock, multiple-edge, state acquisition (logic analyzer)
Connect to
HP 8133A Channel
2 Output
Pod 1, channel 3
Pod 2, channel 3
Pod 3, channel 3
Pod 4, channel 3
Testing Performance
Connect to
HP 8133A Channel
1 Output
J-clock
3–53