HP 1660C Series Service Manual page 81

Logic analyzers
Hide thumbs Also See for 1660C Series:
Table of Contents

Advertisement

Set up the Format menu.
2
a Press the Format key. Select State Acquisition Mode, then select Full Channel/4K
Memory/100MHz.
b Select the field to the right of each pod field, then select ECL. The screen does not
show all pod fields at one time. Use the knob to access pod fields not shown on the
screen.
Set up the Trigger menu.
3
a Press the Trigger key. Select Modify Trigger, select Clear Trigger, then select All in
the pop-up menu.
b Select Count Off. Press Select again, then select Time in the pop-up menu. Select
Done to exit the menu.
c Select the field labeled 1 under the State Sequence Levels. Select the field labeled
"anystate," then select "no state." Select Done to exit the State Sequence Levels menu.
d Select the field next to "a," under the label Lab1. Type the following for your logic
analyzer, then press the Select key.
HP 1660C/CS/CP – "00AA"
HP 1661C/CS/CP – "002A"
To test the single-clock, multiple-edge, state acquisition (logic analyzer)
HP 1662C/CS/CP – "00AA"
HP 1663C/CS/CP – "000A"
Testing Performance