Siemens SITRANS F M MAG 3100 sensor Operating Instructions Manual page 14

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS F M MAG 3100 sensor:
Table of Contents

Advertisement

Safety notes
2.2 Installation in hazardous area
Make sure the hazardous area approval is suitable for the environment in which the device
will be installed.
All approvals are based on non-flammable processes only!
Intrinsically safe data
Table 2- 3
Electrode circuit "ia" (Terminal 82,83)
Ui
Li
Ii
Pi
Ci
Coil circuit "Ex e" (Terminal 85,86)
Ui
Ii
Compact mounted versions
For intrinsically safe data for MAG 3100 Ex compact mounted with MAG 6000 I Ex d, refer
to the Operating Instructions of MAG 6000 I or to certificate number Sira 11ATEX2124X,
available here: Certificates (http://www.siemens.com/processinstrumentation/certificates).
With intrinsically safe circuits, use only certified meters appropriate for the transmitter.
If a non-conforming supply unit is used, the "fail-safe" type of protection will no longer be
effective and the approval certification will be invalid.
12
WARNING
WARNING
Intrinsically safe data for MAG 3100 Ex remote mounted
WARNING
WARNING
30V
2μH
50mA
0.5W
50nF
30 V (70 V peak)
130mA
SITRANS F M MAG 3100 sensor
Operating Instructions, 09/2012, A5E03005599-02