Bluetooth Pairing; Date & Time; Bluetooth; Hanna Lab Application - Hanna Instruments HI9810372 Instruction Manual

Wireless ph tester for skin & scalp
Table of Contents

Advertisement

Bluetooth Pairing

Option: dEL PAIr
Press ON/OFF button to delete saved paired
device.
Date & Time
Option: SEt TIME
Press ON/OFF button to set the date and time.
Option: YEAR, MO, DAY, HOUR, and MIN
Use the Setup button to select the option and press the ON/OFF button to change the selected option.

7. BLUETOOTH

With "
" or "
PAIr BLU
On BLU
tester is in discoverable mode. Once connected, the icon will stop blinking. If not connected, the icon is not displayed.
• Select "
", to enable Bluetooth without bonding.
On BLU
• Select "
", to enable Bluetooth with bonding. A 6-digit bonding pin is displayed the first time the tester and
PAIr BLU
smart device are paired. Once the devices are paired, the bonding pin is not required when reconnected.
• Select "
", to disable Bluetooth.
OFF BLU
• Select "
", to delete all paired devices. If PAIr BLU is enabled, a bonding pin will need to be re-entered.
dEL PAIR

8. HANNA LAB APPLICATION

• The Hanna Lab App is available on the App Store
• Consult the Help section of the application for information on calibration, measurement, data logging, and sharing.
• When the tester is in discoverable mode, it will appear in the list of "Available Devices".
• Within the application, tap "Connect" to pair the tester with the device. All readings are transmitted directly to the
application.

Tag a Measurement

Once connected to the application, the ON/OFF button can be pressed to tag the current reading.
• Press the ON/OFF button from measurement mode. The display shows "
• The reading on the application will flash green and the note icon ( ) will be displayed. Tap the note icon ( ) to
add an annotation.

9. CALIBRATION

Preparation & Guidelines
1. Remove the probe from the plastic storage cap. Save the cap for probe storage.
2. Rinse off any storage solution or salts that may be on the body.
3. Verify there is solution inside the pH bulb by shaking the probe down to restore continuity as the solution may have
moved up the stem during shipping.
4. For best results, use a rinse beaker and a separate calibration beaker for each buffer. Discard rinsing buffers after use.

Procedure

For most applications it is recommended to start with pH 7.01 buffer (or pH 6.86).
To restore factory defaults, press and hold the ON/OFF button. "
Note: It is recommended to calibrate the electrode with buffers at the temperature it will be used at.
App Store is a trademark of Apple, Inc.
Google Play and the Google Play logo are trademarks of Google LLC.
dEL
&
PAIR
2021
YEAR
" selected in Setup, the Bluetooth icon (
®
and on Google Play.
10. 1 2
&
25. 6
CAL CLR
3
MO
) will blink for up to 45 seconds, indicating the
&
1
", followed by "
SEt TAG
-
&
TAG
" is displayed.
15
DAY
".
- TAG
7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Halo 2

Table of Contents