Care and Maintenance
To obtain the highest accuracy for measurements it is important to follow these tips:
• Calibration is only as good as the buffer being used. The pH buffer values change over
time once the sachets are opened. Fresh buffer should be used for each calibration.
• The probe should be rinsed with purified water each time before placing in buffer or
sample to be tested.
• When the meter is not in use it is important to add several drops of storage solution
to the protective cap to keep the probe hydrated. If storage solution is not available,
pH 4.01 or pH 7.01 buffer can be used.
• For improved accuracy it is recommended to calibrate in two buffers.
• It is important to calibrate and measure samples at the same temperature. A dramatic
change in temperature between buffer solutions and samples to be tested will give
.
inaccurate readings
• If fouled, clean the electrode by soaking in cleaning solution for 20 minutes, then
rinse the tip and soak in storage solution at least 30 minutes before use. Recalibrate
after each cleaning.
Warranty
The meter is warranted for a period of one year against defects in workmanship and materials when
used for their intended purpose and maintained according to instructions. This warranty is limited to
repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed
maintenance is not covered. If service is required, contact your local Hanna Instruments Office. If under
warranty, report the model number, date of purchase, serial number and the nature of the problem. If the
repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Returned Goods Authorization (RGA) number from the
Technical Service department and then send it with shipping costs prepaid. When shipping any instrument,
make sure it is properly packaged for complete protection.
Recommendations for Users
Before using this product, make sure it is entirely suitable for your specific application and for the
environment in which it is used. Any variation introduced by the user to the supplied equipment may
degrade the meters' performance. For yours and the meter's safety do not use or store the meter in
hazardous environments.
Certification
All Hanna Instruments conform to the CE European Directives.
Disposal of Electrical & Electronic Equipment. The product should not be treated as household
waste. Instead hand it over to the appropriate collection point for the recycling of electrical and electronic
equipment which will conserve natural resources.
Disposal of waste batteries. This product contains batteries, do not dispose of them
with other household waste. Hand them over to the appropriate collection point for recycling.
Ensuring proper product and battery disposal prevents potential negative consequences for the
environment and human health. For more information, contact your city, your local household waste
disposal service, the place of purchase or go to www.hannainst.com.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of
the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.
Auto-off
From measurement mode, press and hold the
ON/OFF button. The meter will cycle through
"OFF," "CAL," then current auto-off setting.
The default setting is 8 minutes ("d08"). Press
ON/OFF button to change. "d60" is auto-off after
60 minutes, and "d--" disables the auto-off
feature. Press and hold the button to exit the
menu.
Clear Calibration
Place meter in calibration mode. Press and hold
ON/OFF until "CLr" is displayed. The meter will
now be at default calibration.
"Err" Message
In calibration mode, if the meter displays an
"Err" message when in the correct fresh buffer
solution then the probe should be cleaned.
Place the probe in the
solution for 20 minutes. Rinse with purified water
and place in storage solution for a minimum of
30 minutes before calibrating.
Battery Indicator
RoHS
compliant
The meter features a low battery indicator. When
the battery is running low, the tag will blink on
screen. When the battery has been depleted,
"Erb" will appear on screen and the meter will
turn off.
Battery Replacement
To change the CR2032 Li-ion battery, turn the battery cover located on the back of the meter
counterclockwise to unlock. Remove cover and replace the battery with + side facing up.
Note: Batteries should only be replaced in a safe area using the battery
type specified in this instruction manual. Old batteries should be disposed in
accordance with local regulations.
Accessories
pH Buffer Solution
Code
Description
HI70004P
pH 4.01 buffer solution, 20 mL sachets (25 pcs.)
HI70007P
pH 7.01 buffer solution, 20 mL sachets (25 pcs.)
HI77400P
pH 4.01 & 7.01 buffer solution, 20 mL sachets (10 pcs., 5 ea.)
HI700640
cleaning
Electrode Cleaning Solution
Code
Description
HI700601P
General purpose cleaning solution, 20 mL sachets (25 pcs.)
HI700640P
Cleaning solution for milk deposits, 20 mL sachets (25 pcs.)
Cleaning and disinfection solution for dairy products,
HI700641P
20 mL sachets (25 pcs.)
Electrode Storage Solution
Code
Description
HI70300L
Electrode storage solution, 500 mL bottle
HI70300M
Electrode storage solution, 230 mL bottle
HI9072
Electrode storage solution, 13 mL dropper
IST981034 05/18-1
INSTRUCTION MANUAL
HI981034
Milk pH Tester

Advertisement

loading

Summary of Contents for Hanna Instruments Foodcare HI981034

  • Page 1 Damage due to accidents, misuse, tampering or lack of prescribed pH Buffer Solution maintenance is not covered. If service is required, contact your local Hanna Instruments Office. If under “Err” Message warranty, report the model number, date of purchase, serial number and the nature of the problem. If the...
  • Page 2 Preparation: Press the ON/OFF button to turn the meter on. All tags The glass tip uses a special low temperature (LT) glass Thank you for choosing a Hanna Instruments product. Please read this instruction will be displayed. Unstable The pH electrode is shipped with a protective cap containing storage solution. Before formulation.