Securing The Device With Mounting Bracket - Siemens SITRANS P320 Compact Operating Instructions

Pressure transmitter
Table of Contents

Advertisement

4. For insulated systems, ensure that you insulate the device as far as possible to the lower edge
of the enclosure.
In this way, you avoid a defect in the device or the loss of explosion protection for Ex devices.
You can find the permissible temperature values in the section Technical data (Page 63).
5. To guarantee secure and vibration-free installation of the pressure transmitter, fasten it to a
mounting bracket (Page 29).
Level
You can find details on how to mount the device with level in the section Installation (level)
(Page 31).
4.3

Securing the device with mounting bracket

Here you have the following mounting options with the mounting bracket:
• On a mounting range
• On a vertical or horizontal pipe (Ø 50 to 60 mm), according to the examples 1 and 2
Safety notes
NOTICE
Mounting with differential pressure lines
Differential pressure lines can break if they are not mounted correctly.
• Install the device so that the pressure transmitter and the differential pressure lines are not
SITRANS P320/P420
Compact Operating Instructions, 09/2021, A5E38874562-AD
subject to different vibrations.
4.3 Securing the device with mounting bracket
Installing/mounting
29