Agilent Technologies Division TriScroll 600 Series Replacement Manual

Agilent Technologies Division TriScroll 600 Series Replacement Manual

Dry scroll vacuum pump

Advertisement

Quick Links

TriScroll
Dry Scroll
Vacuum Pump
Manual No. 699904305
Revision E
November 2014
600 Series
TM
Vacuum Products Division
MODULE REPLACEMENT
MANUAL

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies Division TriScroll 600 Series

  • Page 1 Vacuum Products Division TriScroll 600 Series MODULE REPLACEMENT MANUAL Dry Scroll Vacuum Pump Manual No. 699904305 Revision E November 2014...
  • Page 2 TriScroll™ 600 Dry Scroll Vacuum Pump TriScroll is a trademark of Agilent. Loctite and PST are registered trademarks of Loctite Corporation. Krytox and Viton are registered trademarks of E. I. du Pont de Nemours and Company. Copyright 2014 Agilent, Inc.
  • Page 3: Table Of Contents

    TriScroll 600 Dry Scroll Vacuum Pump Contents Preface........v Safety Considerations .
  • Page 4 TriScroll 600 Dry Scroll Vacuum Pump This page intentionally left blank.
  • Page 5: Preface

    TriScroll 600 Dry Scroll Vacuum Pump Preface This manual provides the information you need to successfully perform replacement of the entire module of your AgilentTriScroll Dry Vacuum Pump. Module replacement is performed as an alternative to performing a complete overhaul and rebuild (Major Maintenance) and is sold as one of the options recommended when the pump base pressure has risen to an unacceptably high level for your application.
  • Page 6 TriScroll 600 Dry Scroll Vacuum Pump WARNING Disconnect power from the TriScroll 600 before performing any maintenance procedure. Allow the pump to cool before perform- ing any maintenance procedure. Approximate cool-down time is one to two hours. CAUTION Wipe all O-rings clean with a lint-free cloth before installation to ensure that no foreign matter is present to impair the seal.
  • Page 7: Related Triscroll Manuals

    TriScroll 600 Dry Scroll Vacuum Pump Related TriScroll Manuals Manuals related to the installation and operation, tip seal replacement, and major maintenance for the TriScroll 600 series pumps are listed in the following table: Title Applicable TriScroll Model Part Number Major Maintenance Manual All TriScroll 600 Series Models 699904300...
  • Page 8: Factory Service Options

    TriScroll 600 Dry Scroll Vacuum Pump Factory Service Options Agilentoffers factory-rebuild service or advance exchange of complete TriScroll Pumps or TriScroll Pump Modules. Contact your nearest Agilent sales office for price and availability information. Select your preferred service option from the table below. Factory Service Options Part Number Advance Exchange TriScroll 600 Single Phase...
  • Page 9: General Information

    TriScroll 600 Dry Scroll Vacuum Pump General Information AgilentTriScroll 600 series pumps are designed to provide years of trouble-free service if maintenance procedures and intervals are observed. Bearing grease replenishment and tip seal replacement is recommended when the pump base pressure has risen to an unacceptably high level for your application.
  • Page 10: Maintenance Took Kit

    TriScroll 600 Dry Scroll Vacuum Pump Maintenance Took Kit...
  • Page 11 TriScroll 600 Dry Scroll Vacuum Pump Maintenance Took Kit (continued)
  • Page 12: Triscroll Disassembly

    TriScroll 600 Dry Scroll Vacuum Pump TriScroll Disassembly 1. Remove the two M5x16 screws from the intake assembly. 2. Remove the intake assembly from the top of the pump. 3. Remove and discard the O-ring from the groove on the underside of the intake fitting.
  • Page 13 TriScroll 600 Dry Scroll Vacuum Pump TriScroll Disassembly (continued) 4. Remove the three M5x16 screws that attach the cowling to the module. 5. Remove the cowling. 6. Remove the four M6x16 screws that attach the pump module to the frame. 7.
  • Page 14 TriScroll 600 Dry Scroll Vacuum Pump TriScroll Disassembly (continued) 8. Remove M8x12 screw and washer attaching fan assembly to crankshaft. 9. Remove fan assembly from crankshaft. 10. Remove key from slot in crankshaft.
  • Page 15: Triscroll Reassembly

    TriScroll 600 Dry Scroll Vacuum Pump TriScroll Reassembly Tool required: Allen wrench ® Krytox GPL 224 grease Locate the following items shown in the photos on the left: ➀ Pump Module Assembly ➁ Fan Assembly ➂ M6x16 screws (4) ➃ Spider Coupling ➄...
  • Page 16 TriScroll 600 Dry Scroll Vacuum Pump TriScroll Reassembly (continued) 1. Install key in slot on crankshaft. 2. Slide fan assembly onto crankshaft engaging key and against seal spacer. 3. Secure fan assembly to crankshaft with M8x12 screw and washer previously removed in step 8 on page 6.
  • Page 17 TriScroll 600 Dry Scroll Vacuum Pump TriScroll Reassembly (continued) 4. Insert spider in motor coupling. 5. Install pump module on frame, aligning fingers on fan assembly with fingers on coupling. Assure proper fit of dowel pins on mating holes in frame. NOTE Exhaust fitting in the down- ward position.
  • Page 18 TriScroll 600 Dry Scroll Vacuum Pump TriScroll Reassembly (continued) 6. Secure pump module to frame with four M6x16 screws. 7. Install cowling over pump module. Secure with three M5x16 screws.
  • Page 19 TriScroll 600 Dry Scroll Vacuum Pump TriScroll Reassembly (continued) 8. Lightly grease the 2-127 O-ring and insert it into the groove on the intake fitting. 9. Place the intake assembly into the top of the pump. 10. Secure the fitting with two M5x16 screws.
  • Page 20: Pump Conditioning And Performance Verification

    TriScroll 600 Dry Scroll Vacuum Pump TriScroll Reassembly (continued) This figure illustrates a fully reassembled TriScroll 600 Series Pump. Pump Conditioning and Performance Verification Pump Module conditioning and performance verification has been performed at Agilent factory. Further conditioning or performance verification is not necessary.
  • Page 21 Please follow these instructions whenever one of our products needs to be returned. 1) Complete the attached Request for Return form and send it to Agilent Technologies (see below), taking particular care to identify all products that have pumped or been exposed to any toxic or hazardous materials.
  • Page 22 This page intentionally left blank. 4) HEALTH and SAFETY CERTIFICATION AGILENT TECHNOLOGIES CANNOT ACCEPT ANY PRODUCTS CONTAMINATED WITH BIOLOGICAL OR EXPLOSIVE HAZARDS, RADIOACTIVE MATERIAL, OR MERCURY AT ITS FACILITY. Call Agilent Technologies to discuss alternatives if this requirement presents a problem.
  • Page 23 Shipping Regulations (IATA, DOT, etc.) and carrier requirements. Customers receiving an Advance Exchange product agree to return the defective, rebuildable part to Agilent Technologies within 15 business days. Failure to do so, or returning a non-rebuildable part (crashed), will result in an invoice for the non-returned/non-rebuildable part.
  • Page 24 TriScroll 600 Dry Scroll Vacuum Pump This page intentionally left blank.
  • Page 26 Service & Support North America India Southeast Asia Agilent Technologies Agilent Technologies India Pvt. Ltd. Agilent Technologies Sales Sdn Bhd Vacuum Products Division Vacuum Products Division Vacuum Products Division 121 Hartwell Avenue G01. Prime corporate Park, Unit 201, Level 2 uptown 2,...

Table of Contents