Agilent Technologies G6011B User Manual
Agilent Technologies G6011B User Manual

Agilent Technologies G6011B User Manual

Quiet cover ms

Advertisement

Agilent G6011B
Quiet Cover MS
User Manual

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies G6011B

  • Page 1 Agilent G6011B Quiet Cover MS User Manual...
  • Page 2 Notices Warranty Safety Notices © Agilent Technologies, Inc. 2020 No part of this manual may be reproduced in The material contained in this document is any form or by any means (including elec- provided “as is,” and is subject to being...
  • Page 3: Table Of Contents

    Contents Introduction General Information Important Safety Warnings Cleaning Recycling the Product Parts Identification Site Preparation Installation Installing the Oil Tray Accommodating a KF40 Foreline Pump Intake Flange Installing the Rough Pump Inside the Quiet Cover MS Connecting the Power Supply Verifying the Installation Operation Control Panel and Indicators...
  • Page 4: Introduction

    Parts Identification Site Preparation This manual describes how to install, operate, and troubleshoot the Agilent G6011B Quiet Cover MS for LC/MS rough pumps. Your rough pump fits in the Quiet Cover MS to significantly reduce noise, while an internal fan ensures that your pump will not overheat.
  • Page 5: General Information

    The Quiet Cover MS is designed and manufactured under a quality system registered to ISO 9001. Information The Agilent Technologies Quiet Cover MS meets the following IEC classifications: Category II, Pollution Degree 2, Class III. Quiet Cover MS has been designed and tested in accordance with recognized safety standards and is designed for use indoors.
  • Page 6 Warnings in the manual or on the instrument must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
  • Page 7: Important Safety Warnings

    Safety cautions and warnings must be observed during all phases of operation, service, and repair of Quiet Cover MS. Failure to comply with these precautions violates safety standards of design and Quiet Cover MS’s intended use. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
  • Page 8: Cleaning

    • Consult your equipment dealer, Agilent Technologies, or an experienced technician for assistance. • Changes or modifications not expressly approved by Agilent Technologies could void the user’s authority to operate the equipment. Sound emission certification for Federal Republic of Germany •...
  • Page 9: Parts Identification

    Quiet Cover MS kit. Figure 1 Figure 2 on page 10 depict the parts of the Quiet Cover MS. Table 2 G6011B Quiet Cover MS contents list Part Quantity Part number Quiet Cover MS G6011B...
  • Page 10 Parts Identification Parts Identification Fan Intake Kickstand Power connection Figure 2. Quiet Cover MS (rear view) (power supply not shown) Quiet Cover MS User Manual...
  • Page 11: Site Preparation

    Site Preparation Site Preparation Site Preparation Prepare the installation site to meet the criteria in Table Table 3 Site prep information Criteria Description Physical dimensions Weight 11.3 kg (25 lbs) Kickstand up:  Dimensions 68 × 25 × 44 cm (27 × 13.5 × 17.25 inches)  (L ×...
  • Page 12: Installation

    Installation Installing the Oil Tray Accommodating a KF40 Foreline Pump Intake Flange Installing the Rough Pump Inside the Quiet Cover MS Connecting the Power Supply Verifying the Installation This chapter describes how to assemble the Quiet Cover MS and install the rough pump. Quiet Cover MS User Manual...
  • Page 13: Installing The Oil Tray

    Installing the Oil Tray Installing the Oil Tray Installing the Oil Tray Your Quiet Cover MS is shipped to you with the oil tray packaged separately. You need to install it prior to using the Quiet Cover MS. 1 Remove any protective film from the tray. 2 Slide the tray into the bottom of the Quiet Cover MS with the deep part of the tray facing the front of the Quiet Cover MS.
  • Page 14: Accommodating A Kf40 Foreline Pump Intake Flange

    The G6011B Quiet Cover is configured from the factory to accommodate an MS-40+ vacuum pump with a KF25 Foreline Pump intake flange. If you are installing the G6011B Quiet Cover on an MS-40 vacuum pump with a KF25 Foreline Pump intake flange, skip this procedure.
  • Page 15: Installing The Rough Pump Inside The Quiet Cover Ms

    • MS-40S (variable speed electrical pump fan) The G6011B Quiet Cover is configured from the factory to accommodate an MS-40+ vacuum pump. If this revision of pump is being installed, no changes are required and the installer can proceed to the next procedure on...
  • Page 16 Installing the Rough Pump Inside the Quiet Cover MS Installing the Rough Pump Inside the Quiet Cover MS To install the rough pump inside the Quiet Cover MS: 1 Lock the front locking casters to secure the Quiet Cover MS (Figure Main cover Door finger...
  • Page 17 Installing the Rough Pump Inside the Quiet Cover MS Installing the Rough Pump Inside the Quiet Cover MS 3 Install the rough pump into the Quiet Cover MS. Orient the rough pump support foot into the front mounting cup inside the Quiet Cover MS and between the Quiet Cover MS mounting tabs so that the pump oil level window sight glass faces the front access door (Figure Motor air duct...
  • Page 18: Connecting The Power Supply

    Connecting the Power Supply Connecting the Power Supply 7 Install the main cover assembly. Confirm the cover stud guide pins are aligned properly with the main chassis assembly. 8 Close the front access door. 9 Connect MS Foreline hose to the pump. Refer to pump documentation or more information. 10 If the Quiet Cover MS is not currently in its operating location, unlock the locking casters, carefully push the Quiet Cover MS to its operating location, and lock the front locking casters.
  • Page 19: Verifying The Installation

    Verifying the Installation Verifying the Installation Verifying the Installation When the G6011B Quiet Cover is plugged into an external power outlet, the Power Switch (front NOTE lower left-hand corner of the Quiet Cover) will light up orange. This indicates that power is connected to the Quiet Cover.
  • Page 20: Operation

    Operation Control Panel and Indicators Handling Temperature Alarms Changing Rough Pump Oil Using the Kickstand To operate Quiet Cover MS, you will need to: • Understand the controls and indicators on the front panel • Know how to handle temperature alarms and signals •...
  • Page 21: Control Panel And Indicators

    Control Panel and Indicators Control Panel and Indicators Control Panel and Indicators Figure 10 shows the location of the on/off switch and indicators of the Quiet Cover MS control panel. Status light Power switch Figure 10. Control panel Handling Temperature Alarms In normal operation the Quiet Cover MS fan will turn off and on to maintain the temperature set limits.
  • Page 22: Changing Rough Pump Oil

    Changing Rough Pump Oil Changing Rough Pump Oil Changing Rough Pump Oil Periodically, you must change the rough pump oil. The oil tray of the Quiet Cover MS is designed to collect the oil as it drains from the pump. Please refer to your pump documentation for detailed information regarding changing your pump oil.
  • Page 23 Changing Rough Pump Oil Changing Rough Pump Oil 10 Reinstall the drain plug when the oil has completely drained. 11 Remove the oil tray and dispose of the oil according to local safety and environmental guidelines. 12 Return the Quiet Cover MS to its untilted position. See “Using the Kickstand”...
  • Page 24: Using The Kickstand

    Using the Kickstand Using the Kickstand Using the Kickstand To tilt the Quiet Cover MS: 1 Make sure the Quiet Cover MS is on a level surface. 2 Lock the two locking casters located on the front of the Quiet Cover MS (Figure 14).
  • Page 25 Using the Kickstand Using the Kickstand 4 Step down on the horizontal bar and lift gently on the back end of the Quiet Cover MS while stepping on the kickstand until the Quiet Cover MS is in its tilted position (Figure 16).
  • Page 26: Troubleshooting

    Troubleshooting Table 4 describes solutions to problems that may occur with Quiet Cover MS. If your problem is not listed or the solution listed does not work, consult your pump documentation or contact your Agilent service representative. Table 4 Problems and solutions Problem Solution Pump does not fit in Quiet Cover MS.
  • Page 27  Agilent Technologies, Inc. 2020 Second edition, April 2020 *G6011-90110* G6011-90110...