Installation and Setting Instructions
With the laser light localizer
Fig. 1:
Remove the laser light localizer prior to the installation procedure. After installation is
complete, remount the localizer, check it, and adjust it as needed according to the "Laser
Light Localizer" setting instructions.
Installing the measuring chambers
•
Disconnect the ARCADIS power plug.
•
Unscrew the POWERPHOS cover. Carefully separate the adhesive joints on the long
sides of the cover.
•
Remove the measuring chamber connection
•
Place the dose measuring chamber on the POWERPHOS according to
•
Use the included threaded bolts and screws to secure the dose measuring chamber to
the POWERPHOS.
•
Connect the measuring chamber connection
•
Use the switch on the measuring chamber to select the desired resolution (see the in-
scription on the measuring chamber and the measuring chamber operating instruc-
tions).
•
Reattach the POWERPHOS cover.
Siemens AG
Medical Solutions
Measuring chamber connection
SPR2-330.814.01.01.02
02.05
(1/Fig. 1 / p. 5)
from the cable tie.
(1/Fig. 1 / p. 5)
to the measuring chamber.
Page 5 of 10
CS SD 24
5
0
0
(2/Fig. 1 / p.
5).
ARCADIS Avantic