Option area dose measurement
Hide thumbs Also See for ARCADIS Avantic:
Table of Contents

Advertisement

Installation and Setting Instructions
Fig. 2:

Checking the programmed value (standard resolution/pediatrics)

NOTE
Click <Diamentor> and ">" under "Adjustment" for the next page in the Service menu.
Click "Click Here" in the service view and check the accuracy of the diamentor accord-
ing to the instructions on the system.
Siemens AG
Medical Solutions
main_system_
After completing all work, ensure that the value entered under
"Constant" and "Resolution" corresponds with the application de-
sired by the user.
"0.01cGy cm*cm/pulse" is only to be selected for pediatrics.
SPR2-330.814.01.01.02
02.05
Page 7 of 10