Installation
Alternative Instrument Mounting Procedure
When it is not convenient to work from the back of the panel, the instrument can be connected
before fitting in the panel.
1. Install the panel support frame in the panel opening
2. Slip the cables through the panel opening
3. Remove the instrument cover
4. Slip the cables through the instrument cover
5. Slip the cables through the instrument back panel cable glands
6. Connect the cables to the instrument electronic boards
7. Tighten the cable glands
8. Reinstall the instrument cover
9. Install the instrument in the panel opening
3.4.3 Connections (bottom of instrument)
1 - Power cable
2 - Ethernet cable gland
3 - USB-A host connector for mass storage device
4 - USB-B 4 pin connector (reserved for
manufacturing)
Note: The type of power cable connection will vary depending on the instrument specification.
26
Figure 11 Panel mount connection panel
5 - Input/Output 1 cable gland
6 - Sensor connection
7 - External pressure sensor connection
8 - Input/Output 2 cable gland