Residual Parameters - HP Q METER 4342A Operating And Service Manual

Table of Contents

Advertisement

Model 4342A
Parallel
Connection
Measurements
Section
III
Paragraphs
3-71 and 3-72
3-71.
Dielectric
Measurement.
3-72.
The dielectric
constant
and dielectric
loss
of insulating
materials
can be measured
by a method
similar
to and is basically
a
capacitance
measurement.
When a pair
of
parallel
electrodes
(air
capacitor)
connected
to 4342A (in
air)
and an insulating
material
placed
between
the
electrodes,
the electrode
capacitance
increases
in proportional
to the
specific
inductive
capacity
(ES) of the
sample material.
The dielectric
constant
of
the sample material
is calculated
as the pro-
duct
of Es and the vacuum dielectric
constant
CO.
Accordingly,
the dielectric
constant
can
be determined
from the
capacitance
measure-
ments made before
and after
placing
the
sample between
the elecrodes.
Additionally,
after
the sample
is mounted
in the holder,
the
conductance
of the sample
can also
be
calculated
from a reduction
of the Q meter
indication.
To make easy and accurate
dielectric
measurements,
it
is recommended
that
the
16451A Dielectric
Test Adapter
be
used with
the
4342A.
Typical
characteristics
of the 16451A are described
in Table
3-2.
Materials
to be measured
with
the 16451A
should
be less
than
1Omm in thickness
and
from 38 to 55nn in diameter.
When measuring
materials
with
a high
dielectric
constant
or
a large
loss,
it
is usually
best
to prepare
material
in thicknesses
greater
than
3nn.
On the other
hand,
when low loss material
is
to be measured,
the material
thickness
should
be less
than
3mm.
Materials
measuring
less
than
0.5mm in thickness
are usually
difficult
to measure.
To make dielectric
measurements
using
the
16451A,
proceed
as follows:
a.
b.
CO
d.
e.
f.
Depress
the appropriate
FREQUENCY
RANGE button
and set FREQUENCY dial
control
for
the desired
measurement
frequency.
Select
a reference
inductor
which
can
resonate
at the measurement
frequency.
Connect
it
to 4342A measurement
COIL
(HI and LO) terminals.
Adjust
L/C dial
and AC dial
controls
for
a maximum Q meter
deflection.
Note sum of the C dial
and AC dial
readings
as Cl and panel
meter
read-
ing as Qi.
Let
the reference
inductor
remain
in
place
(as is)
and attach
the
16451A to
4342A measurement
CAPACITOR (HI and
GND) terminals.
Set 16451A electrode
spacing
as de-
sired.
However,
if
possible,
it
is
best
to set the electrode
spacing
dimension
to about
the
same as the
thickness
of the material
to be
measured.
Again resonate
the measurement
circuit
by adjusting
the L/C and AC dial
con-
trols.
Note C dial
and AC dial
read-
ings
as C2 and panel
meter
reading
Qa.
Table
3-2.
16451A (4342A-KOl)
Typical
Characteristics.
Electrode
Diameter:
38mm
Electrode
Spacing:
0 ~1Omm variable
Minimum vernier
division:
0.02nm
Residual
Parameters:
Co=5pF
Go<0.4uS
(at
1OMHz)
Loz40nH
Minimum measurable
loss
angle
(tan
6):
Approximately
1 x 10e4
3-21