HP Q METER 4342A Operating And Service Manual page 59

Table of Contents

Advertisement

Model 4342A
Table
5-6.
Adjustable
Components.
Section
V
Table
5-6
Reference
Designator
I
Name of Control
I
Purpose
AlAlCl,
-c3,
-c5,
-C7,
-c9,
-Cll,
-Cl3
FREQ. ADJ C
To adjust
oscillator
frequency
of upper
range
limits
of individual
frequency
ranges
to
maximize
frequency
accuracies.
AlAlLl
thru
AlAlL
7
FREQ. ADJ L
To adjust
oscillator
frequency
of lower
range
limits
of individual
frequency
ranges
to
maximize
frequency
accuracies.
HIGH FREQ. ADJ
To adjust
Q meter
sensitivity
in high
frequency
(7OMHz) region
to obtain
optimum
frequency
flatness.
ASR32
MED FREQ. ADJ
To adjust
Q meter
sensitivity
in high
frequency
(2OMHz) region
to obtain
optimum
frequency
flatness.
ASR43
LINEARITY ADJ
To adjust
Q meter
linearity
to maximize
measurement
accuracy.
A6R2
A6R4
I::
A6R6
Xl GAIN
To set Q meter
full
scale
sensitivity
to
maximize
measurement
accuracy.
REC BAL
To set Q ANALOG OUTPUT dc voltage
at rear
panel
to zero volts
at Q meter
zero scale
deflection.
REC GAIN
To set Q ANALOG OUTPUT dc voltage
at rear
panel
(to
1 volt)
at Q meter
full
scale
deflection.
A6R21
X10 GAIN
To adjust
AQ measurement
full
scale
sensitivity
to maximize
AQ measurement
accuracy.
A7R3
Q-PRESET
To properly
set Q limit
selector
sensitivity
at
Q LIMIT
dial
full
scale
setting
to maximize
the
dial
scale
accuracy.
A7R7
Q-PRESET
VERNIER
To properly
set Q limit
selector
sensitivity
at
Q LIMIT
dial
center
scale
setting
to maximize
the dial
scale
accuracy.
A8R12
-RZl
+25V ADJ
-25V ADJ
To set +25V and -2SV output
voltages
of dc
power supply.
A8R26
I
OSC LEVEL
I
To adjust
oscillator
output
voltage
to maximize
measurement
accuracy.
5-7