Page 2
If there is a failure of the fuse cap with one contact terminals, replace the fuse cartridge with cap with a new version. The new fuse cartridge (fuse cartridge) has two contact terminals with a bayonet design. Components Affected - VIDEOMED DI / DIM / DI-K Required Materials - Fuse block 30 60 761...
Page 3
Insert the new melt insert into the fuse cartridge. • Insert the fuse cartridge with the melt insert into the fuse block. Siemens AG Register 5 RX52-022.091.01 Page 3 of 4 VIDEOMED DI / DIM / DI-K Medical Engineering Rev. 02 07.97 TD AX 4...
Page 4
Page 4: "Final Work Steps" changed. "Changes to Previous Version" added. TD AX 4 / Link TD AX 1 / Biedermann / Tropia VIDEOMED DI / DIM / DI-K Register 5 RX52-022.091.01 Page 4 of 4 Siemens AG Rev. 02 07.97...