Repair Instructions
Procedure
Observe ESD guidelines!
Switch the generator OFF.
Switch the system breaker OFF.
Preparatory Worksteps
Disconnect all cables from the TV CCU.
Remove the front shielding plate from the TV CCU.
Remove the transformer unit;
to this, only loosen the four mounting screws and pull out the unit.
Remove the TV CCU.
Remove the rear EMI shield.
Removing the Old Fuse Block
Remove the solder from the 4 connector terminals for the F8 fuse block on the backplane
board.
Use the solder removal tool!
Remove the fuse block from the backplane.
Completely remove the solder from the soldering terminals with the solder removal tool.
Install the New Fuse Block
Insert the new fuse block turned 45° into the adjacent diagonal holes.
Solder in the connector terminals.
Insert the new melt insert into the fuse cartridge.
Insert the fuse cartridge with the melt insert into the fuse block.