Siemens VIDEOMED DI Repair Instructions page 2

Advertisement

2
General Remarks
The transfer resistance between the fuse cartridge and the holder can cause excess heat-
ing and to a failure.
In an error situation, Error 007 (AP-ID 16) is displayed;
If there is a failure of the fuse cap with one contact terminals, replace the fuse cartridge
with cap with a new version.
The new fuse cartridge (fuse cartridge) has two contact terminals with a bayonet design.
Components Affected
- VIDEOMED DI / DIM / DI-K
Required Materials
- Fuse block
(with fuse cartridge)
Spare Parts
- Melt fuse insert, 8 A
Required Tools
- Standard service tool kit
- Soldering iron, 60 W
- Solder removal tool
- Allen wrenches,
M3 and M4
VIDEOMED DI / DIM / DI-K
with the Lithostar Multiline:9020 on the control PC;
with the Uroskop D1:Error 007 (AP-ID 10)
Register 5
30 60 761
77 51 142
20 49 708 RH 090
RX52-022.091.01
Rev. 02
07.97
Repair Instructions
Page 2 of 4
TD AX 4
Medical Engineering