Final Work Steps
Reinstall the rear EMI shield.
Insert the TV CCU in the cabinet and secure it in place.
Insert the transformer unit and tighten the mounting screws.
Install the front shielding plate.
Connect all cables.
Function Check
Switch the system breaker ON.
Switch the generator ON.
After the system has been initialized, switch on fluoroscopy without an object in the beam
path.
At the generator, a tube voltage of less than 45 kV must stabilize when Automatic Fluoro
Control is switched on.
The fluoro image will appear on the monitor.
Switch fluoroscopy OFF.