4
Final Work Steps
•
Reinstall the rear EMI shield.
•
Insert the TV CCU in the cabinet and secure it in place.
•
Insert the transformer unit and tighten the mounting screws.
•
Install the front shielding plate.
•
Connect all cables.
Function Check
•
Switch the system breaker ON.
•
Switch the generator ON.
•
After the system has been initialized, switch on fluoroscopy without an object in the beam
path.
•
At the generator, a tube voltage of less than 45 kV must stabilize when Automatic Fluoro
Control is switched on.
•
The fluoro image will appear on the monitor.
•
Switch fluoroscopy OFF.
Changes to Previous Version
Page 1:
Page 2:
Page 4:
TD AX 4 / Link
TD AX 1 / Biedermann / Tropia
VIDEOMED DI / DIM / DI-K
Updated.
Paragraph "General Remarks": Lithostar and Uroskop added.
Paragraph "Procedure" changed.
"Final Work Steps" changed.
"Changes to Previous Version" added.
Register 5
RX52-022.091.01
Rev. 02
Repair Instructions
Page 4 of 4
07.97
TD AX 4
1
1
1
Siemens AG
Medical Engineering