Mounting and sensor wiring
February 2023
Procedure
1. Remove the transmitter-to-sensor wiring compartment cover to reveal the terminal
2. Feed the sensor wiring cable into the transmitter wiring compartment.
3. Connect the sensor wires to the appropriate terminals.
20
connections.
Figure 3-8: Removal of the transmitter-to-sensor wiring compartment cover
Figure 3-9: Sensor wiring feedthrough
Important
Terminate the four-wire cable drain wires only at the sensor/core processor end of
the cable. See the sensor installation manual for more detail. Do not connect the
four-wire cable drain wires to the ground screw located inside the junction box.
See
Figure 3-10
for four-wire terminal connections.