Vna Main Menu Keys; Calibrate Configure Menu - HP 85097A User Manual

Electronic calibration system
Table of Contents

Advertisement

Operating an ECal System

VNA Main Menu Keys

VNA Main Menu Keys
After selecting the
menu keys appear on the network analyzer display.
Calibrate Configure Menu
This key accesses the menu for selecting the calibration configuration
you want.
1. Configure the calibration using the following softkeys.
INTERPOLATE < ON I OFF >
Select
interpolation capability of the ECal system.
• With interpolation on, you can measure any frequency within the
• With interpolation off, only frequencies that correspond to
ISOLATION AVERAGES
Select
the VNA keypad. Values from 1 to 4096 are valid for the HP 8510 VNA.
Values from 1 to 999 are valid for the HP 87xx VNA series.
The speed of the calibration is determined by the number of isolation
NOTE
averages.
For all VNA calibration measurements (other than isolation), you
perform measurements with the averaging value specified in the VNA
setup. You may want to increase the number of averages significantly
for isolation measurements.
2. Once completed, select
2-14
Start
INTERPOLATE: ON/OFF
ISOLATION AVERAGES
CONFIDENCE CHECK
SERVICE FUNCTIONS
RETURN
INTERPOLATE < ON I OFF >
span of the characterized modules.
characterization frequencies of the modules are allowed.
ISOLATION AVERAGES
button of the ECal software, the following
to enable or disable the frequency
and set the desired number of averages on
to return to the ECal menu.
RETURN
Chapter 2