Download Print this page

Emerson Rosemount MultiVariable 4088A Quick Start Manual page 30

With modbus output protocol

Advertisement

Quick Start Guide
Markings
I6 CSA Intrinsically Safe
Certificate 2618446
Standards CSA C22.2 No. 0-10, CSA C22.2 No. 25-1966, CSA C22.2 No.
Markings
8.6
Europe
E1 ATEX Flameproof
Certificate FM12ATEX0030X
Standards EN 60079-0:2012+A11:2013, EN 60079-1:2014, EN
Markings
Special Conditions for Safe Use (X):
1. This device contains a thin wall diaphragm less than 1 mm thickness
that forms a boundary between Category 1 (process connection) and
Category 2 (all other parts of the equipment). The model code and
datasheet are to be consulted for details of the diaphragm material.
Installation, maintenance and use shall take into account the
environmental conditions to which the diaphragm will be subjected.
The manufacturer's instructions for installation and maintenance
shall be followed in detail to assure safety during its expected
lifetime.
2. Flameproof joints are not intended for repair.
3. Appropriate cable, glands, and plugs need to be suitable for a
temperature of 5 °C greater than the maximum specified
temperature for location where installed.
4. The applicable temperature class, ambient temperature range and
process temperature range of the equipment is as follows:
30
Class I, Division 1, Groups B, C, D; Class II, Division 1, Groups E,
F, G; Class III; Class I, Division 2, Groups A, B, C, D; Temp Code
T5; seal not required; when installed per Rosemount Drawing
04088-1053; Type 4X
30-M1986, CSA C22.2 No. 94-M91, CSA C22.2 No. 142-
M1987, CSA C22.2 No. 157-92, CSA C22.2 No. 213-M1987,
CSA C22.2 No. 60079-0:2011, CSA C22.2 No. 60079-11:2011,
ANSI/ISA 12.27.01-2003
Class I, Division 1, Groups C, D, Temp Code T3C; Class I Zone 0
Ex ia IIB T4; when installed per Rosemount Drawing
04088-1207; Type 4X
60079-26:2015, EN 60529:1991+A1:2000
II 1/2 G Ex db IIC T6...T4 Ga/Gb, T4/T5(–50 ≤ T
50 ≤ T
≤ 65 °C)
a
Rosemount 4088A MultiVariable Transmitter Quick Start Guide
April 2019
≤ 80 °C), T6(–
a

Advertisement

loading

This manual is also suitable for:

Rosemount 4088a