Download Print this page

Configuration Screens - Siemens Healthineers UniPOC RAPIDPoint 500 Configuration Manual

Advertisement

RAPIDPoint 500/RAPIDPoint 500e Configuration Guide

4 Configuration Screens

The Instrument Configuration screens in UniPOC allow you to view and edit the RAPIDPoint 500 and
RAPIDPoint 500e configuration data for a specific location. These devices can be configured for the entire
Organization or for an individual Facility, Department or Location.
Note: Siemens recommends entering configuration data at the Facility level when possible. Department
level settings should then be used to vary the instrument configurations only as needed. The Organization
level settings then remain available for reference as default values.
Procedure:
Select a Facility, Department or Location in the tree.
From the menu, select Instruments > Configuration.
The Instrument Configuration screen appears.
Select RAPIDPoint 500 from the Instrument drop-down list.
Select one of the following Configuration Download options:
Assigned Location
Download Location
Select the desired configuration tab and set the options.
Note: Be sure to note the selected Tree level prior to saving the instrument configuration
settings to ensure the settings are applied to the appropriate locations.
Do one of the following:
Click the Save Configuration button to save the settings.
This will break the Parent-Child relationship and save the configuration as a new one, even
if no changes were made. Saving an instrument configuration for a lower level of the Tree
(Department or Location) will sever the relationship between the child and its parent level on
RAPIDPoint 500/RAPIDPoint 500e Configuration Guide
(Recommended) If you select this option, UniPOC uses the Location
where the device's serial number or name resides in the Location Tree
regardless of where the device is used or docked (Default). This location
will determine the:
Configuration settings to be downloaded to the device
Supported lists downloaded to the device
Location where all test data are uploaded to in the Tree
(Not Recommended) If you select this option, UniPOC uses the Location
where the device's or docking station's IP address resides in the Location
Tree. This location will determine the:
Configuration settings to be downloaded to the device
Supported lists downloaded to the device
Location where all test data are uploaded to in the Tree
Note: UniPOC will automatically update the device's location in the Tree
when it communicates to UniPOC.