Printing Measurement Data; Disconnecting From The Printer - Konica Minolta BC-10 Plus Instruction Manual

Baking meter
Table of Contents

Advertisement

Connecting a Printer

Printing Measurement Data

This function can be used to print measurement data.
The instrument must be connected to the printer in advance. See page 63 "Connecting
a Printer" for details on connecting the instrument to a printer.
Notes
Printing is not possible if the connection is not correct.
• Do not disconnect the cable while printing. Failure to do so may cause a
malfunction.
Setting Procedure
Connect the instrument to the printer and enable
communication.
Press  "Back" from the <MENU> screen to go to the
measurement screen.
Perform measurement.
The measurement results will be displayed on the instrument's LCD screen
and at the same time the printer will begin printing.
Print Sample
Serial No.
S/N 3000002 BC-10 Plus
01 BCU3.30
Instrument name display
L* 56.7 a* 6.2 b* 26.4
or
S/N 3000002 BC-10 Plus
03 BCU3.32
When the screen display is switched
S/N 3000002 BC-10 Plus
01 L* 56.7 a* 6.2 b* 26.4
When data is corrected
S/N 3000002 BC-10 Plus
Delete
(Cont.)

Disconnecting from the Printer

10:00
01
3.70
BCU
63.8
L*
7.4
a*
Notes
29.3
b*
When continuous measurement is performed
S/N 3000002 BC-10 Plus
[First measurement]
01 BCU3.26
L* 56.1 a* 6.2 b* 26.2
S/N 3000002 BC-10 Plus
[Second measurement]
02 BCU3.26
L* 56.1 a* 6.2 b* 26.2
[Third measurement]
S/N 3000002 BC-10 Plus
03 BCU3.26
L* 56.1 a* 6.2 b* 26.2
Average display
S/N 3000002 BC-10 Plus
Average of three measurements
BCU Avg3.26 Max3.26 Min3.26
Disconnect the printer from the instrument.
MENU/OK
Press
, move the selection frame to "USB
Connec." using either
or
, and then press
MENU/OK
.
Move the selection frame to "Disconnect" using either
MENU/OK
or
, and then press
To print again, turn the printer OFF once, and
reconnect the printer from the fi rst step.
Click  "Back" to go back to the <MENU> screen.
S/N 3000002 BC-10 Plus
Avg03 L* 56.1 a*
6.2 b* 26.2
S/N 3000002 BC-10 Plus
Max01 L* 56.1 a*
6.2 b* 26.2
S/N 3000002 BC-10 Plus
Max03 L* 56.1 a*
6.2 b* 26.2
S/N 3000002 BC-10 Plus
BCU Avg3.26 Max3.26 Min3.26
AVERAGE
Ver.1.10.0000
MENU
Back
Avg. Reset
Browse Data
Mask Setting
Auto Off
Date/Time
USB Connec.
10:00
MENU
.
Back
USB Connec.
Disconnect
PC
Printer
When display format is switched
L*a*b* average values
AVERAGE
Maximum values
Minimum values
Average, maximum, and
minimum BCU values
64

Advertisement

Table of Contents
loading

Table of Contents