Examination procedure
Radiation time and dose display
(1) Display of accumulated fluoroscopic time
at the C-arm system
(2) Display of accumulated fluoroscopic time
at the monitor trolley (Examination task
card)
(3) Display of accumulated dose at the moni-
tor trolley (with optional dose measuring
chamber)
Alternative: Display of air kerma values
(preceding item: accumulated air kerma).
After every five minutes of radiation, an
audible alarm sounds at the C-arm sys-
tem.
To temporarily deactivate the warning sig-
nal, press the Reset fluoro time button
on the C-arm system (4).
ADR stop, manual input of radiation
parameters
All exposures are taken with automatic expo-
sure control. This rule is turned off using
ADR stop. The kV values can then be
selected manually.
Select ADR stop at (5).
If required, set a manual value with the
+/- buttons (6).
kV and mA values are shown on the dis-
play at the C-arm system (7).
As long as ADR stop is activated, the
manually set value remains unchanged.
27