Advertisement

5. While shorting across the CAL keypad on the PCA, turn the rotary switch to the function to
adjust. See
Figure
6. When CAL shows on the display, remove the short.
This puts the Product into calibration mode.
7. When the real value shows on the display, apply the correct input signals from
the Product. For each function, the calibration step shown on the display will advance.
8. After each step, push
to the next step.
9. Set the Calibrator to Standby after you complete adjustment of each function.
If any calibration point is missing, Err is shown on the display.
After you push
the calibrator source. Some adjustment steps can take several seconds to execute
before the Product goes to the next step.
10. When calibration adjustment is complete, remove the 3.0 V supply.
The Product exits the calibration mode automatically.
11. Replace the batteries and battery door.
CAL Keypad
5.
on the front to confirm the calibration step, store the value, and go
, wait until the calibration step number advances before you change
Figure 5. CAL Keyboard
Note
Clamp Meter
Calibration Adjustment
Table 5
to
Battery Contacts
13