Fluke 301A Calibration Manual page 13

Clamp meter
Hide thumbs Also See for 301A:
Table of Contents

Advertisement

Frequency
Value
Unit
Amplitude
A ac (with 50-turn coil)
0.2
0.8
2
8
A
50 Hz
12
14
19.5
A ac, Hz (with 50-turn coil)
45.0
0.2 A
400.0
Hz
45.0
8 A
400.0
A dc (with 50-turn coil) must zero for each point
-16
-6
-0.6
0.2
1.2
A
0 Hz
2
12
14
19.5
Table 3. Current Performance Tests
or
Res.
Spec.
0.01
0.01
0.1
0.1
2 %
0.1
1
1
0.1
0.1 %
1
0.1
0.01
0.01
0.01
2 %
0.1
0.1
1
1
301 Model
Counts
A
B
C
• • • • •
10
• • • • •
10
• • • • •
5
5
• • • •
5
5
5
• • • • •
3
• • • • •
3
• • • • •
3
• • • • •
3
5
5
10
10
10
5
5
5
5
Clamp Meter
Performance Tests
Lower
Upper
Limit
Limit
D
E
9.7
10.3
39.10
40.90
97.5
102.5
391.5
408.5
587.5
612.5
681
719
951
1000
44.7
45.3
399.3
400.7
44.7
45.3
399.3
400.7
-821
-779
• •
-306.5
-293.5
• •
-30.70
-29.30
• •
9.70
10.30
• •
58.7
61.3
• •
97.5
102.5
• •
587.5
612.5
681
719
951
1000
9

Advertisement

Table of Contents
loading

This manual is also suitable for:

301b301c301d301e

Table of Contents