5 - 6
Work sequence:
1.
2.
3.
4.
5.
6.
7.
Average deviation
Basic position
C-arm to the left
C-arm to the right
ARCADIS Avantic
Select full size (no zoom).
Adjust the Iris diaphragm to an opening of about 15 cm.
Adjust the C-arm (see Fig. 3).
Switch on fluoroscopy briefly.
Read the evaluation at the TV Monitor (Fig. 4).
(Use stored image).
Calculate the deviation
2
2
x
+ y
:
Test results (see also Fig. 4):
(equal to > 80% over-
lapping of the visible
(equal to > 80% over-
lapping of the visible
(equal to > 80% over-
lapping of the visible
permissible*
must be < 47 mm
area)
must be < 47 mm
area)
must be < 47 mm
area)
SPR2-330.662.01
Rev. 02
05.06
Function tests
Average deviation
...................... mm
...................... mm
...................... mm
Page 6 of 6