IRECTION
Chapter 3 Environmental (HVAC)
requirements
Section 3.1 Relative humidity and temperature
Storage values refer only to equipment that is still in shipping containers. If the equipment is
partially or completely installed, refer to in-use values.
Ensure that the in-use and storage locations meet the temperature and humidity minimums and
maximums listed in
Table 3-1 Environmental requirements - relative humidity and temperature
Ensure that the in-use and storage location conditions do not fluctuate more than the maximum
rates of change given in
Table 3-2 Maximum rates of change for relative humidity and temperature
Section 3.2 Altitude
Storage values refer only to equipment that is still in shipping containers. If the equipment is
partially or completely installed, refer to in-use values.
Ensure that the in-use and storage locations meet the altitude minimums and maximums listed in
Table
Table 3-3 Altitude requirements, relative to sea level
3
EVISION
Table
3-1.
Relative humidity (non-condensing)
In-use
Min.
Max.
Min.
30%
75%
25%
Table
In-use
Temperature change
< 10° C (50° F)/hour
Humidity change
3-3.
In-use
Min.
-30 m
3,000 m
(-98.4 ft)
(9843 ft)
Chapter 3 Environmental (HVAC) requirements
Storage
Max.
Min.
10° C
85%
(50° F)
3-2.
Storage
< 20° C (68° F)/hour
< 30%/hour
Max.
Min.
-30 m
(-98.4 ft)
O
XR200
PTIMA
AMX
P
-I
RE
NSTALLATION
Temperature
In-use
Storage
Max.
Min.
35° C
-10° C
(95° F)
(14° F)
< 30%/hour
Storage
Max.
3,000 m
(9843 ft)
/XR220
AMX
M
ANUAL
Max.
45° C
(113° F)