SELFTEST PROCEDURES

DISPLAY TEST
94
The selftest procedure can be made only entering the setup
menu and selecting one of the items of the "tESt" group
(t .00 - t .08) .
Note
All the tests are made while in setup mode, where a time-out
is present . If no action is performed for about 5 minutes,
the mode is automatically exited and the instruments returns
to previous mode .
The display selftest procedure consists of lighting up all the
display segments together .
•To start the procedure select
setup item t .00 and an "OFF"
blinking message will appear on
the primary LCD .
•Press the  (or ) key once and
the message will switch to a
blinking "GO" .
•Press CFM key to confirm or the
 (or ) key again to return to
the previous status .
•If confirmation is given when
the "OFF" message is blinking,
no action is performed and it
will move to the next setup item
(t .01) .
•The display test is announced by a scrolling "Display test"
message .
• All the segments light up for a few
seconds and then switch off before
exiting the display test procedure
and moving to the next setup item
(t .01) .