Table 3
Gas Sample Selector specifications
Feature
General
Sample Input Pressure/Purge
Pressure
Repeatability
Size
Weight
Electrical supply
Storage Environmental
Requirements With 8890/8860
Operating Environmental
Requirements
Automatic System Purging
Sample Points
Injection Position Indication
Communication with GC
Control with Software
Control w/o Software
26
Cylinder Version
Cylinder Version Gas Sample Selector
Input pressure: 25 bar (363 psi) max
Purge gas pressure: 25 bar (363 psi) for both
versions
Actuator gas pressure: 3.8 bar (55 psi)
8890: ≤1.5%
8860: ≤2%
Note: Using 8890/8860 with PCM, and Agilent
Data System for analysis of 5% methane in
Argon. Results may vary with other samples
and conditions.
H 255 mm, W 284 mm, D 426 mm.
11.6 KG w/o package
24 Vdc, 90 W Max
Humidity: 5% to 95% RH (noncondensing)
Storage extremes: ~40 to 70 °C
Humidity: 5% to 90% (noncondensing)
Temperature: 15 to 35 °C
Operating altitude 4,600 m above sea level
Different purge modes are available.
Purge time and flow can be adjusted.
8 Max
LED
USB
Sequence control fully supported by
8860/8890.
Firmware 2.2 and Driver 3.4 or later version is
required.
8890/8860 Local UI/GC Browser UI
Agilent Gas Sample Selector for 8860 and 8890 User Manual
Bag Version
Bag Version Gas Sample Selector
Input pressure: 0.2 bar (2.9 psi) max
Purge gas pressure: 25 bar (363 psi) for both
versions
Actuator gas pressure: 3.8 bar (55 psi)
8890: ≤1.5%
8860: ≤2%
Note: Using 8890/8860 with PCM, and Agilent
Data System for analysis of 5% methane in
Argon. Results may vary with other samples
and conditions.
H 257 mm, W 190 mm, D 426 mm.
8.3 KG w/o package
24 Vdc, 90 W Max
Humidity: 5% to 95% RH (noncondensing)
TStorage extremes: ~40 to 70 °C
Humidity: 5% to 90% (noncondensing)
Temperature: 15 to 35 °C
Operating altitude 4,600 m above sea level
Different purge modes are available.
Purge time and flow can be adjusted.
8 Max
LED
USB
Sequence control fully supported by
8860/8890.
Firmware 2.2 and Driver 3.4 or later version is
required.
8890/8860 Local UI/ GC Browser UI