AGFA ADC Solo Maintenance Manual page 17

Hide thumbs Also See for ADC Solo:
Table of Contents

Advertisement

DD+DIS013.02E
Edition 1, Revision 10
(2) Expose a new image plate:
Place the cassette in length
direction to the X-ray tube,
see figure below.
Set the following exposure
parameters:
7.5 mAs, 77 kVp, 1.3 m
distance
Doses 10 µGy (result of
setting: 7.5 mAs, 77 kVp, 1.3
m distance)
1.5 mm Cu-filter with small
focus
Turn cassette by 180°.
Expose plate a second time
by using the same
parameters.
Notice, that all exposure
parameters are approximate
values.
(3) Identify the cassette on the ID Station:
In the <Patient name> field, type a name and a cassette format,
e.g. Flatfield 18 x 24.
In the <First name> field, type the serial number of the digitizer,
e.g. SN1356.
In the <Birth date> field, type the current date,
e.g. 20012001 (use date format xxyyzzzz for day/month/year).
In the <Radiologist> list, click <SERVICE>.
In the <Examination> list, click <system diagnosis>.
In the <Sub-examination> list, click <Flat field>.
Confirm the Exposure class <200>.
Make sure that the outlined areas are filled in as shown in the
example.
(Type 5155 / 100/200)
Rotating
Anode
ADC Solo
Maintenance
x-ray tube
Cathode
Figure 10
Section 12 / 13