Laser camera timing (not for MEMOSKOP 2)

Timing diagram

This timing diagram is for reference only.
Monitor B, BAS
Hardcopy, BAS
TV = Precontact time default = 100 ms
TK = Contact time, default = 200 ms
TN = Postcontact time default = 5000

Changing the times

Notice

Prerequisites

Input can only be done if there is a MEMOSKOP keyboard on hand.
The values will only be displayed if there is a text module. However, "blind" programming
can also be done.
For UROSKOP and Multiline, plug D1.X6 on the MEMOSKOP must be removed when
programming the times.

Opening Setup

Press the [NEXT PAT] key.
Next, press the [@] and [%] keys in sequence .
Press the [CTRL] + [S] keys simultaneously.
A list of numbers will be displayed on the monitor.

Changing the precontact time TV

Enter the following : 00006506
**
means in HEX code: 01 = 100 ms, 02 = 200 ms, 32 = 5000 ms.
Then press the [CTRL] + [E] keys simultaneously.
Then, depending on the requirement, program the contact time TK and/or postcontact
time TN or exit the Hardcopy Setup (refer to the section "Exit Setup ").
Siemens AG
Medical Engineering
, Image ON
Image OFF
, Image ON
Image OFF
Relay contact ON
Relay contact OFF
Adhere to the manufacturer's instructions for the particular
camera being used.
RX57-029.061.01
Rev. 02
TV
TK
**
Page 1 of 4
10.97
TD SD 24
7 - 1
7
TN
MEMOSKOP 2/3/50/100/SUB
7
7
7
7