Uroskop D1/D2; Activate Set-Up Mode; Exit Set-Up Mode - Siemens MEMOSKOP 2 SUB Service Instructions Manual

Table of Contents

Advertisement

4 - 2

UROSKOP D1/D2

Switch the system "ON"
Wait for the MEMOSKOP to boot.

Activate set-up mode

NOTICE
Press the following keys on the keyboard in sequence:
- <NEXT PAT>
- <@>
- <%>
- <CTRL> + <S>
After activating set-up mode, a sequence of numbers is displayed in the lower image
region of the monitor.
To call up the test images, overwrite this sequence of numbers.
Enter the number sequence without any blanks.
Depending on the test image, reconstruction may take up to 30 seconds.
Test images:
- SMPTE-similar test image
- Gray wedge horizontal
- Black/white step
- Gray scale 0%-100%(linear)
- Gray scale 0%-100% (exponential)
- Black 0%
- Vertical white bar 60%
Press keys <CTRL> + <E>.

Exit set-up mode

Switch the system "OFF" and back "ON" again.
The memory returns to normal operating mode.
MEMOSKOP 2/3/50/100/SUB
<x> means: press key x.
<CTRL> + <E> means: press and hold <CTRL> and simulta-
neously press key <E>.
RX57-029.061.01
Rev. 03
Calling up the service menu
0000 8808 0A0303C F00
0000 8808 0A03021 C00
0000 8808 0A03023 C00
0000 8808 0A03025 C00
0000 8808 0A03026 C00
0000 8808 0A03027 C00
0000 8808 0A03028 C00
Page 2 of 2