Measure Temperature

The black extenders (long and short) of the BTL21 Interactive Test Probe each has an
IR temperature sensor. The IR temperature sensors can measure the temperature of
battery poles (negative pole).
To measure the temperature of a battery pole:
1. Aim the IR sensor on the battery pole.
2. Adjust the test lead to let the entire flare fall on the battery pole. See Figure 4-5.
The temperature reading shows on the LCD of the interactive test probe.
3. Push the Save key on the interactive test probe.
The temperature reading is saved in the Product memory. The reading can only be
accessed through a PC.
Figure 4-5 shows the measurement field of a battery pole.
BTL21
Test Lead
sales@GlobalTestSupply.com
Figure 4-5. Measure the Temperature of a Battery Pole
To set objective surface emissivity in accordance with battery post
material, push Setup > Handle > Emissivity. See the Emissivity
section.
Fluke-Direct
Use the BTL21 Interactive Test Probe
Note
.com
Measure Temperature
D = 11.5 mm
hsz023.eps
1.888.610.7664