Specifications; General Specifications - Fluke BT21ANG User Manual

Battery analyzer
Table of Contents

Advertisement

Specifications

General Specifications

 Fuse Protection for Resistance ...................... 0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse, Fluke specified part
Power Supply
Battery power ...................................................... BP500 smart battery pack: double cell lithium-ion, 7.4 V, 3000 mAh
Battery life ........................................................... >8 hours in continuous full-load operation
Battery charging time .......................................... ≤4 hours
Power adapter output voltage ............................. Use only BC500 battery charger: 18 V, 840 mA
Line power .......................................................... 100 V ac to 240 V ac adapter with country specific plug
Frequency ........................................................... 50 Hz to 60 Hz
Temperature
Operating ............................................................ 0 °C to 40 °C
Storage ............................................................... -20 °C to 50 °C
Lithium-ion battery charging ............................... 0 °C to 40 °C
Relative Humidity (non-condensing, 10 °C)
Operating ............................................................ ≤80 % at 10 °C to 30 °C
Altitude
Operating ............................................................ 2,000 m
Storage ............................................................... 12,000 m
Temperature Coefficient ....................................... 0.1 x (specified accuracy) /°C (<18 °C or >28 °C)
Size ......................................................................... 58 x 103 x 220 (mm)
Weight .................................................................... 850 g
Memory
Data/Setup flash memory ................................... 4 MB
Real-Time Clock .................................................... Time and date stamp for measurement. The RTC works >50 days
IP Rating ................................................................ IEC 60529: IP 40
Safety ..................................................................... IEC 61010-1, IEC 61010-2-030, IEC 61010-031, Pollution Degree 2
EMI, RFI, EMC, RF ................................................. IEC 61326-1, IEC 61326-2-2, EN 300 328, EN 301 489-1, EN 301 489-
Electromagnetic Compatibility ............................ Applies to use in Korea only. Class A Equipment (Industrial
[1]
This product meets requirements for industrial (Class A) electromagnetic wave equipment and seller or user should take notice of it.
This equipment is intended for use in business environments and is not to be used in homes.
sales@GlobalTestSupply.com
only
≤75 % at 30 °C to 40 °C
without battery.
600 V CAT III, 1000 V dc max; Derated to CAT II with CAT II probe cap
installed
17, FCC Part 15 Subpart C Sections 15.207, 15.209, 15.249
CONTAINS FCC IDs: T68-FWCS, XDULE40-S2
IC: 6627A-FWCS, 8456A-LE4S2
Broadcasting & Communication Equipment)
Fluke-Direct
Product Overview and Specifications
[1]
.com
1.888.610.7664
1
Specifications
1-11

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bt521

Table of Contents