This measurement example will help you measure the absolute phase noise
of a microwave source (2.5 to 18 GHz) with frequency drift of ≤10E – 9 X
Carrier Frequency over a period of thirty minutes.
To prevent damage to the Agilent/HP 70420A test set's hardware
components, the input signal must not be applied to the signal input
connector until the input attenuator has been correctly set for the desired
configuration, as show in
when the Connection Diagram appears.
The following equipment is required for this example in addition the phase
noise test system and your unit-under-test (UUT).
To ensure accurate measurements, you should allow the UUT and
measurement equipment to warm up at least one hour before making the
noise measurement.
Table 7-17
Equipment
Agilent/HP 8644B
Agilent/HP 70422A
Coax Cables
Agilent Technologies E5500 Phase Noise Measurement System 7-97
Absolute Measurement Examples
Table 7-19 on page
7-103. Apply the input signal
Required Equipment for the Microwave
Source Measurement Example
Quantity
Comments
1
Must have DCFM Input Port.
Refer to
Reference"
about reference source
requirements
1
Must be entered in the Asset
Manager and Server Hardware
Connections dialog box.
And adequate adapters to connect
the UUT and reference source to
the test set.
Microwave Source
Chapter 6, "Selecting a
for more information