Troubleshooting/Faqs; Sensor Check - Siemens SITRANS FM MAG 3100 Operating Instructions Manual

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS FM MAG 3100:
Table of Contents

Advertisement

Troubleshooting/FAQs

7.1

Sensor check

Requirement
To check the SITRANS FM sensors the following test instruments will be required:
• Digital Meter/Multimeter
• Megger
• (Moving Coil Meter)
Sensor check
Remove the transmitter from the sensor or remote position before making the following checks.
Coil resistance check
• Measure the coil resistance between connection numbers 85 and 86 using a digital meter.
Resistance should be within range stated in Coil resistance table (Page 59).
A low reading may indicate moisture within the coil housing or shorted coil tums.
A high reading would indicate an open circuit coil.
Note
In case of deviation from nominal coil values, the sensor is damaged and must be replaced
Coil insulation check
Potential hazard!
Only carry our a coil insulation check in non-hazardous area!
• Megger between connection number 85 and the sensor body.
The resistance should be above 20 MΩ.
A low megger reading would indicate the coil insulation is breaking down. This is normally due
to fluid ingress into the coil housing.
Sensors with an insulation resistance down to 1 MΩ may still work satisfactorily but this is not
guaranteed.
SITRANS FM MAG 3100
Operating Instructions, 12/2021, A5E03005599-AE
WARNING
7
55

Advertisement

Table of Contents
loading

Table of Contents