Connect the common test lead before the live test lead and
remove the live test lead before the common test lead.
Remove all probes, test leads, and accessories before the
battery door is opened.
Do not use and disable the Product if it is damaged.
Do not use the Product if it operates incorrectly.
Do not use test leads if they are damaged. Examine the test
leads for damaged insulation, exposed metal. Check test
lead continuity.
Before each use, examine the Product. Look for cracks or
missing pieces of the clamp housing. Also look for loose or
weakened components. Carefully examine the insulation
around the jaws. See Figure 1, item .
Examine the case before you use the Product. Look for
cracks or missing plastic. Carefully look at the insulation
around the terminals.
Read all safety information before you use the Product.
Remove batteries to prevent battery leakage and damage to
the Product if it is not used for an extended period.
Remove the batteries if the Product is not used for an
extended period of time, or if stored in temperatures above
50 °C. If the batteries are not removed, battery leakage can
damage the Product.
To prevent possible damage to the Product or to equipment
under test, use a thermocouple rated for the temperatures to be
measured. The Product is rated for -10.0 °C to +400.0 °C and
14 °F to 752 °F. The included type-K thermocouple is rated to
260 °C.
Figure 1. Clamp Jaw (325 is Shown)
Caution
2
CAT
600 V
CAT
300 V
1
Clamp Meter
Safety Information
hfk002.eps
3