Fluke 323 Calibration Manual page 17

Clamp meter
Hide thumbs Also See for 323:

Advertisement

Test
(Switch Position)
AC Volts
DC Volts
Ohms
AC Amps
(with 50-Turn Coil)
Temperature
AC Volts
DC Volts
Ohms
[1] At each function, after all calibration points is adjusted, the display shows "Save" after the last
calibration point. Push HOLD to save the calibration data.
Table 5. Calibration Adjustment
LCD Reading
323
C-1.1
C-1.2
C-1.3
C-1.4
C-2.2
C-3.1
C-3.2
C-4.1
C-4.2
C-4.3
C-4.4
[1]
324 and 325
C-21
C-22
C-23
C-00
C-01
C-02
C-03
C-04
C-05
C-06
C-07
C-08
C-09
C-10
C-11
C-12
C-13 (325 only)
C-14 (325 only)
Calibration Adjustment
Calibrator Output
1.05 V, 50 Hz
40 V, 50 Hz
10 V, 50 Hz
300 V, 50 Hz
400 V
400 Ω
4000 Ω
0.019 A, 50 Hz
0.42 A, 50 Hz
1.026 A, 50 Hz
6 A, 50 Hz
0 V
0.02 V
0.02 V
600 V, 50 Hz
600 V, 100 Hz
600 V, 200 Hz
600 V, 300 Hz
600 V, 400 Hz
0 V
600 V
0 V
0.5 V
0 Ω
400 Ω
440 Ω
4000 Ω
4400 Ω
40000 Ω
Clamp Meter