323/324/325
Users Manual
Remove batteries to prevent battery leakage and damage to the Product
if it is not used for an extended period.
Remove batteries to prevent battery leakage and damage to the Product
if is to be stored above its operating temperature.
Do not use a current measurement as an indication that a circuit is safe
to touch. A voltage measurement is necessary to know if a circuit is
hazardous.
To avoid possible damage to the Product or to equipment under test, use a
thermocouple rated for the temperatures to be measured. The Product is
rated for -10.0 °C to +400.0 °C and 14 °F to 752 °F. The included type-K
thermocouple is rated to 260 °C.