OMRON Oxygen Concentrator
Y-506W
Instruction manual
*
Thank you for purchasing the OMRON oxygen concentrator.
In order to use the device safely, please read the complete manual carefully before using it for the first time.
Keep this instruction manual in a convenient place for future reference.

Advertisement

Table of Contents
loading

Summary of Contents for Omron Y-506W

  • Page 1 Y-506W Instruction manual Thank you for purchasing the OMRON oxygen concentrator. In order to use the device safely, please read the complete manual carefully before using it for the first time. Keep this instruction manual in a convenient place for future reference.
  • Page 3: Table Of Contents

    CONTENT Ⅰ. Intended use Ⅱ. Special warning Ⅲ. Instructions to use Ⅳ. Safety instructions Ⅴ. Product introduction Ⅵ. Product use environment and normal working conditions Ⅶ. Transportation and storage conditions Ⅷ. The scope of application of the product Ⅸ. Product features Ⅹ.
  • Page 4: Ⅰ. Intended Use

    Ⅰ. INTENDED USE Oxygen inhalation in the general home and relief of respiratory disease. Oxygen in the air is concentrated to support the respiration of patients living at home with respiratory problems, mainly chronic obstructive pulmonary disease, etc. Ⅱ. SPECIAL WARNING 1.
  • Page 5 improper operation . 8. Do not use the oxygen concentrator if either the plug or power cord is damaged . Do not use extension cords or electrical adapters. 9. Do not attempt to clean the concentrator while it is plugged into an electrical outlet. 10.
  • Page 6: Ⅴ. Product Introduction

    No open flames Ⅴ. PRODUCT INTRODUCTION The Y-506W oxygen concentrator uses the principle of Pressure Swing Adsorption to separate the oxygen in the air from nitrogen and other gases. At room temperature, the oxygen which meets medical standards can be continuously separated from the air when the power is turned on.
  • Page 7: Ⅸ. Product Features

    Ⅸ. PRODUCT FEATURES ◆ Accumulative timing function: The total working time of the oxygen concentrator is displayed on the display screen. ◆ Single-time timing function: The single-time working time is displayed on the display screen, understand the running time of the equipment better. ◆...
  • Page 8: Ⅹ. Know Your Device

    Maximum outlet pressure: Under the rated oxygen flow, 20kPa~70kPa 250kPa ± 50kPa Compressor relief valve release pressure: Maximum atomization rate: ≥0.2mL/min Oxygen concentrator noise: ≤60dB(A) Rated input power: 480VA Electrical classification: Class II equipment, type BF applied part Operation mode: Continuous operation Degree of protection against IPX0...
  • Page 9 Humidification bottle Nebulization outlet Atomizer Primary filter Primary filtration Power switch Device power jack (Including fuse) Rear view of the whole device Secondary filter Side view of the whole device...
  • Page 10 1. Operation panel: Time unit: h Time unit: min Indicator light Display screen Button Display screen: It can display the device’s cumulative running time, single-time running time, adjustable timing time, fault code, oxygen purity. a) After power on, the screen displays the accumulated time. The running time starts to count, and the oxygen purity is monitored and displayed in real time.
  • Page 11 Button: Voice: Control device voice function on or off; Timing +: Increase the timing time. Each time you press it, increase 10 mins; Timing --: Decrease the timing time. Each time you press it, decrease 10 mins; ON/OFF: Control device to start or to stop oxygen production; The user needs to inhale oxygen regularly, and adjust the timing time through the "Timing +"...
  • Page 12 6. Fuse holder: When the electric current of the device is overload, the fuse will automatically cut off the Power switch circuit to protect the device and the user. Please replace a new fuse. Fuse Holder 7. Power switch: Device power jack "...
  • Page 13: Ⅺ. Operation Steps

    Ⅺ. OPERATION STEPS Humidification Pressure Oxygen Oxygen 1. Add water to the humidification bottle: Take out tube relief valve inlet outlet the humidification bottle, unscrew the lid, pour distilled water or cold boiled water into the cup, between the maximum and minimum. Then tighten the lid, and put it back to its original position.
  • Page 14 9. If a sudden power failure occurs during the operation of the device with an alarm sound, please check whether the power connection part or the external power supply is in good condition. 10. Nebulization function description: 1) Unscrew the plug of the nebulization outlet and connect it to the disposable kit for nebulization to start nebulization.
  • Page 15: Ⅻ. Maintenance Instructions

    "E4", the red light is on, and the sound indicates. ◆ Under normal conditions, if the product has a low oxygen purity alarm or prompt, please turn it off immediately, use spare oxygen, and immediately contact the OMRON customer support for consultation (see the back cover) .
  • Page 16 extend the oxygen concentrator life. Please clean or replace in time. Primary filter: Remove the cover of the primary filter, take out the filter sponge to clean it once every half a month. Use light detergent firstly, rinse with clean water, and then install it after it is completely dry.
  • Page 17: Schematic Diagram Of Gas Circuit And Electric Circuit

    pry out the fuse holder (Figure 1 and Figure 2). After replacing the new fuse (Figure 3), reinstall the fuse holder (Figure 4). Fuse Figure 1 Figure 2 Figure 3 Figure 4 XIII. SCHEMATIC DIAGRAM OF GAS CIRCUIT AND ELECTRIC CIRCUIT 1.
  • Page 18: Troubleshooting

    Fuse Compressor ~220V50Hz Main control board Combination valve Note: If maintenance needed, please contact the Omron customer support (see the back cover). XIV. TROUBLESHOOTING Failure phenomenon Possible reason Elimination method 1) Poor contact between the power cord plug and the socket.
  • Page 19 The oxygen concentrator 1) Stop using the device immediately, does not work, the 1) The compressor circuit is open. please contact the customer support red light is on, and the for consultation (see the back cover). display shows "E1". The oxygen concentrator 1) Stop using the device immediately, does not work, the 1) The compressor circuit is short.
  • Page 20: Componet Spare Parts And Accessories

    Note: The component spare parts and accessories need to follow "The required transportation and storage conditions."(Please refer to P3). Otherwise the shelf life might be inaccurate. If you need other accessories, please contact the Omron customer support (see the back cover).
  • Page 21: Environmental Information

    The related wastes, residues and accessories of this product may have an impact on the environ- ment, and their disposal should comply with the corresponding national laws and regulations. It is recommended to consult the environmental protection department or the Omron customer support when dealing with the corresponding waste and residue generated from this device.
  • Page 22: Warning

    field or electromagnetic interference source. XVIII. ELECTROMAGNETIC COMPATIBILITY PART DESCRIPTION This product (Hereinafter "This Product" refers to the oxygen concentrator Y-506W) complies with IEC 60601-1-2 "Electromagnetic Compatibility of Medical Devices." The following examples should be followed: 1.This product should be used according to the electromagnetic compatibility information provided in the manual;...
  • Page 23 Table 17.1 Guidance and Manufacturer's Declaration-Electromagnetic Emissions The device is intended for use in the electromagnetic environment specified below.The customer or the user of the device should assure that the device is used in such an environment. Electromagnetic Emissions Test Compliance Environment-Guidance The device uses RF energy only...
  • Page 24 Table 17.2 Guidelines and Manufacturer's Declaration-Electromagnetic Immunity The concentrator is intended for use in the electromagnetic environment specified below. The user of the concentrator should make sure it is used in such an environment. Electromagnetic Environment- IEC 60601 Compliance Immunity Test Test Level Level -Guidance...
  • Page 25 Table 17.3 Guidelines and Manufacturer's Declaration-Electromagnetic Immunity The concentrator is intended for use in the electromagnetic environment specified below. The user of the concentrator should make sure it is used in such an environment. IEC 60601 Compliance Immunity Test Electromagnetic Environment--Guidance Test Level Level Portable and mobile RF communications equipment...
  • Page 26 Table 17.4 Recommended Separation Distances between Portable and Mobile RF Communications Equipment and This Device: This concentrator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of the concentrator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment(trans- mitters)and this concentrator as recommended below, according to the maximum output power of the communications equipment.
  • Page 28 Failure caused by wrong behavior which is not based on instruction manual; f ) Failure caused by improper repair by a repair shop not authorized by Omron, etc. 3. The maintenance service outside the warranty scope will be charged according to the regulations (door-to-door fee, replacement parts fee, etc.).