8.2
Erasure Conveyor (Cleaning/Replacing the Erasure Lamps)
1Y
2Y
CAUTION
Since the erasure lamp may be heated to a high temperature, exercise care not to suffer burns.
(1) Remove the five erasure lamps.
REFERENCE
To remove the erasure lamp, hold the end (metal portion) of the erasure lamp, and initially
snap the left side of the erasure lamp out of the socket.
[Remove] Erasure lamp (x5)
009-058-03
08.30.2002 FM3476
At intervals of one year, or when 60,000 IPs have been processed
¡ Clean.
At intervals of two years, or when 120,000 IPs have been processed
¡ Replace.
#2
Socket
Erasure lamp assembly
CR-IR347 Service Manual
#1
[Remove] BR4x8 (x2)
PM - 46
Bracket
FR7H6040.EPS
PM - 46