Fluke 2063513 Instruction Sheet

Fluke 2063513 Instruction Sheet

Multimeter with temperature

Advertisement

Multimeter with Temperature
P
Read First: Safety Information
To ensure that the meter is used safely, follow these instructions:
Do not use the meter if the meter or test leads appear damaged, or if
you suspect that the meter is not operating properly.
Disconnect the live test lead before disconnecting the common test
lead.
When using the probes, keep your fingers behind the finger guards on
the probes.
Do not use the V Check mode to measure voltages in circuits that
could be damaged by this mode's low input impedance
( 2 k ).
Turn off power to the circuit under test before cutting, desoldering, or
breaking the circuit. Small amounts of current can be dangerous.
Do not apply more than 600V rms between a meter terminal and earth
ground.
Use caution when working with voltages above 60V dc or 30V ac rms.
Such voltages pose a shock hazard.
PN 2063513 December 2002 Rev.1, 9/04
2002,2004 Fluke Corporation. All rights reserved. Printed in China
Instruction Sheet
®
16

Advertisement

Table of Contents
loading

Summary of Contents for Fluke 2063513

  • Page 1 Do not apply more than 600V rms between a meter terminal and earth ground. Use caution when working with voltages above 60V dc or 30V ac rms. Such voltages pose a shock hazard. PN 2063513 December 2002 Rev.1, 9/04 2002,2004 Fluke Corporation. All rights reserved. Printed in China...
  • Page 2: Automatic Range Selection

    Symbols Press button. Press button to switch between modes. Double insulation. Manual ranging mode. ip14i.eps Automatic Range Selection The meter defaults to autoranging when you turn it on. The 4000 mV range can be entered only with manual range selection. Manual Range Selection MULTIMETER To return to autoranging, press...
  • Page 3: Ac And Dc Voltage

    AC and DC Voltage Also refer to V Check. Volts AC Volts DC Input Impedance 5 M Input Impedance 10 M 50 Hz to 400 Hz MULTIMETER MULTIMETER 4000 mV 4V 40V 400V 600V AC / DC AC / DC 600V 600V ip01i.eps...
  • Page 4 W V Check If a dc or ac voltage greater than about 4.5V is present across the inputs R, G, or e when the meter is set to , the meter switches automatically to dc or ac voltage mode (V Check mode). Warning Repetitive transients on a dc bus will cause V Check to select ac volts, even though a hazardous dc voltage may...
  • Page 5 Temperature MULTIMETER Check • TEMPERATURE 600V Vent 80BK Integrated Pipe Temperature Probe (Type K) ip19f.eps Note correct connector polarity. To meet stated accuracy, the 80BK temperature adapter must be at the same temperature as the meter. Warning To avoid possible electric shock, DO NOT apply thermocouple tip to any conductor that is greater than 30V AC, 42.4V pk, or 60V DC to earth.
  • Page 6 Capacitance E Turn off circuit power; then disconnect and discharge the capacitor before measuring capacitance. MULTIMETER If the capacitor requires more discharging, diSC is displayed while the capacitor discharges. Check • TEMPERATURE 600V 10 F 100 F 10,000 F ip05i.eps Note correct probe polarity for polarized capacitors.
  • Page 7 Microamps A Range 0 to 200 A MULTIMETER µA AC µA DC Check • TEMPERATURE 600V Flame Control sensor module probe ip20i.eps Example: Flame rectification circuit. To measure flame rectification circuits: Turn function switch to far right position. Push select button 3 times to select A. Connect meter between flame sensor probe and control module.
  • Page 8: Diode Test G

    Diode Test G Turn off circuit power before testing. For best results diodes should be measured out of circuit. Also refer to V Check. Good Diode Typical voltage MULTIMETER MULTIMETER for silicon diode Low Impedance V•Check Check Check • • TEMPERATURE TEMPERATURE 600V...
  • Page 9 MIN MAX (Records the lowest and highest measurements) V Check , autoranging, and Battery Saver are disabled. Put the meter in the proper range before entering MIN MAX. When the reading changes more than about 50 digits, the meter gives a short beep.
  • Page 10: Min Max With Elapsed Time

    MIN MAX with Elapsed Time Records the hours and minutes between when MIN MAX was entered and the last high and low was recorded. OL is displayed for times longer than 99:59. To enable the MIN MAX timer, hold down while turning the rotary switch from OFF to either measurement mode.
  • Page 11: Disabling The Beeper

    Disabling the Beeper To disable the beeper for all modes, hold down for 2 seconds while turning the meter on. Continuity Capture  To set up the meter to capture intermittent shorts and opens, turn the switch to , connect the leads to the circuit; then press ip12i.eps Captures transitions longer than 250 s (1/4000 of a second).
  • Page 12: Maintenance

    This meter should be serviced only by a qualified service technician. To locate an authorized service center, call: USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-675-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-446-5500 Or, visit Fluke's Web site at www.fluke.com.
  • Page 13: Specifications

    Specifications Accuracy is specified for a period of one year after calibration, at 18 C to 28 C (64 F to 82 F) with relative humidity to 90%. AC conversions are ac-coupled, average responding, and calibrated to the rms value of a sine wave input.
  • Page 14 Function Range Resolution Accuracy Burden Voltage 0 to 200 A 0.1 A (2% + 3 counts) <5 mV/ A (50 Hz to 400 Hz) 0 to 200 A 0.1 A (1% + 2 counts) <5 mV/ A Overload Function Input Impedance (Nominal) Protection >5 M <100 pF 600V rms...
  • Page 15 MIN MAX Recording with Elapsed Time Elapsed Time Resolution Accuracy 0 to 100 hours (99:59) 1 minute 0.3% typical Maximum Voltage Between any Terminal 600V rms and Earth Ground: Display: 3 3/4-digits, 4000 counts, updates 4/sec Operating Temperature: -10 C to 50 C (14 F to 122 F) Storage Temperature: -30 C to 60 C (-22 F to 140 F) indefinitely (to -40 C (-40 F) for 100 hrs)
  • Page 16 Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.