4.4

Troubleshooting

4.4.1

Temperature error (item EM)

This item helps you configuring correctly by showing the most recent temperature error encountered by the option
board.
xx00
No error
xx50
Element 1 detected absent
xx51
Element 2 detected absent
xx52
Element 3 detected absent
xx53
Element 4 detected absent
xx54
Element 5 detected absent
xx55
Element 6 detected absent
xx56
Element 7 detected absent
xx57
Element 8 detected absent
xx58
Element 9 detected absent
xx59
Element 10 detected absent
xx60
Element 11 detected absent
xx61
Element 12 detected absent
xx62
Element 13 detected absent
xx63
Element 14 detected absent
xx = 22 for MPU emulation
xx = 24 for HPU emulation
xx = 30 for HCU emulation
4.4.2

Temperature status request (item MQ)

Four bytes status information (B0 B1 B2 B3) with the following meaning:
For decoding of the status bytes, refer to the ASCII table in appendix A.
Status byte 0:
Contains the characters '0' ÷ 'D', indicating the highest submersed spot of the MPT element. At start up this is: 'I'.
Status byte 1:
Bit 0: general temperature fail
1: fail in average product temperature
2: fail in average gas temperature
3: level exceeds lowest spot element
4: level exceeds highest spot element
5: spot element fail (one or more spots defect)
6: 1
7: 0
Status byte 3:
Bit 0: no previous store command
1: alternative (lower) element selected
2÷5: 0
6: 1
7: 0
Note:
Only the bits which are set to '1' have an active status.
Instruction manual VITO
Average temperature measurement with MPT / RTD
xx70
Sub-system error. Check or replace VITO
xx72
Not supported element type. Check item MT
xx73
Wrong element configuration. Check item MN
xx80
R test error (out of limits)
xx81
R cable error (out of limits)
xx84
Standard deviation too high
xx91
Wrong J0 .. JD setting. Check these items.
xx92
No data available. Check wiring or connection with VITO
Status byte 2:
Bit 0: last valid level used
1: manual level used
2: 0
3: 0
4: 0
5: 0
6: 1