5.2.3 Wiring External Power Disconnection Switches
The mains power must be connected through one or more external 2-pole disconnection switches, so that the
power to the BioTector and/or to the optional air compressor can be isolated without opening the electronics
enclosure.
The external power disconnection switches must be located in easily accessible locations, with a
•
maximum distance of 2 meters from the BioTector and/or the air compressor.
•
The switches must be clearly marked for their purpose.
The switch must comply with local electrical regulations, and have a breaking capacity of 10 Amps or
•
greater.
The MCB or fuse (circuit breaker) located on the customers fuse board should be rated at 10 Amps or
•
greater.
Figure 10 below illustrates the positioning and the installation of the disconnection switch.
Figure 10 External Power Disconnection Switch