4. Replace electronic cover.
Make sure that no cables get pinched.
5. Tighten screws of electronic cover with
1.2 Nm (± 0.5 Nm).
6. Set up backrest to seat.
7. Check all the functions of the bath lift.
WARNING!
– The hand control should only be replaced as
a complete component.
– Do not open or attempt to repair the hand
control.
Please contact Invacare customer service by phone
or order a new hand control using the electronic
spare parts catalog at www.invacare.eu.com.

4.4 Replacing Suction Cups

Difficulty:
●○○
Pos.
Designation
A
Pair of suction cups
8
Tool: —
Article
Quantity
number
15071
2
Fitting
1.
Put rear suction cup with threaded bolt A on sleeve
nut B and screw it in.
Screw as far as the suction cup fits the orientation
on the graphic.
2.
Put front suction cup with threaded bolt A on sleeve
nut B and screw it in.
Screw as far as the suction cup fits the orientation
on the graphic.
3. Insert toggle C in corresponding holding fixture D.
Removal
1. Removal takes place in reverse order.
Fig. 4-3
A
Fig. 4-4
1650139-A