4 Repair

Repairs require a high level of expertise. These assembly
instructions therefore break down the various tasks into
3 categories:
Requirement
Easy – technical understanding required
Medium – technical knowledge required
Difficult – technical knowledge and expertise
in assembling bath lifts required
The required tools and their sizes are listed before the
instructions.
Tool
Phillips head screwdriver
Slot head screwdriver
Torx screwdriver
Snap ring pliers
WARNING!
– To avoid the risk of contamination, clean and
disinfect the bath lift before carrying out
repairs.
IMPORTANT!
– Repairs may only be carried out by an
authorized dealer.
– If possible, continue to use the old
identification label; if this is not possible,
the new identification label must contain
the same information and the old serial
number. (Replacement of spare parts with
serial numbers).
– When components are replaced it is necessary
to ensure the traceability of the components
replaced.
– If screws with thread locker are loosened,
these must be replaced with new ones.
– If screws with circlip rings are loosened, these
must be replaced with new ones.
– Parts that become damaged during removal
must be replaced with new ones.

4.1 Spare Parts

WARNING!
Original spare parts must be used in all repairs.
Otherwise the warranty and product declaration
of conformity shall be rendered invalid.
All spare parts must be obtained from the Invacare
customer service department. An electronic spare parts
catalogue can be found at www.invacare.eu.com.

4.2 Replacing Side Flaps

●○○
Difficulty:
1650139-A
Symbol
●○○
●●○
●●●
Symbol
Removal
1. Press in the clips of side flap A and slowly draw out
Fitting
1. Slide the clips of side flap A evenly into the slots of

4.3 Replacing Hand Control

Difficulty:
Removal
1. Remove battery.
2. Remove backrest from seat and put backrest on
3. Remove screws of electronic cover.
4. Carefully lift up electronic cover from electronic box.
5. Remove sealing.
6. Disconnect and remove old hand control from circuit
Fitting
1. Put black grommet A of hand control into intended
2. Insert plug C of new hand control into connector D
Tool: —
3. Insert new sealing.
Pos.
Designation
Side flap
A
Fig. 4-1
the side flap.
the seat plate B until they engage.
●●●
Pos.
Designation
Hand control
A
incl. sealing
worktop.
board.
D
Fig. 4-2
gap B of electronic box.
on circuit board.
Repair
Article
Quantity
number
1563646
2
A
B
Tool:
Article
Quantity
number
1590349
1
C
A