Dr Calibration Tool Stalls - AGFA XD 10 Series User Manual

Detector calibration
Table of Contents

Advertisement

40 | DR Detector Calibration | Problem solving

DR Calibration Tool Stalls

Problem
Cause
Workaround
Problem
Cause
Workaround
0134I EN 20210924 1204
The DR Calibration Tool stalls.
A virus scanner interferes with the DR Cali-
bration Tool.
Disable the virus scanners on the workstation
and restart the calibration procedure.
The DR Calibration Tool stalls for a short
time.
Information must be retrieved from the DR
Detector.
Wait until the communication to the DR De-
tector has finished and the DR Calibration
Tool reacts again. This may take a number of
seconds.