Replacing the Type-N Output Connector

Tools Required:
9/16 inch open-end wrench
Refer to
Figure
7-9.
1. Remove the connector body with the open-end wrench.
2. Remove the male center conductor assembly from the gold plated inner connector. Use your fingers
to pull out the assembly. The small insulator may or may not come out attached to the center
conductor assembly. If it does, remove it from the old center conductor assembly and place it on the
new one. If it is not attached to the center conductor assembly (it remained inside the inner
connector), proceed to step 3.
3. Hold the new center conductor assembly by the large round insulator, not by the gold plated center
conductor. Carefully align the new center conductor and insert it into the inner connector. Press
gently until you feel it click into place.
4. Place the new connector body and tighten with the open-end wrench.
Figure 7-9
Output Connector Parts
NOTE
Refer to
Chapter 7
"Replaceable Parts" on page 6-4
Replacement Procedure
for part numbers.
Service
7-11