GUIDANCE AND MANUFACTURER'S DECLARATION —
ELECTROMAGNETIC IMMUNITY: The goLITE BLU is intended for use in
the electromagnetic environment specified below. The customer or the
user of the goLITE BLU should make sure it is used in such an environment.
i
mmuniTy
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical Fast
Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
4222.002.7504.1.indd 26
T
iEC 60601 T
EsT
l
EvEl
±(2,4,6) kV contact
±(2,4,8) kV air
±2 kV for power
supply lines
±1 kV for input-
output lines
±1 kV differential
mode
±2 kV common
mode
<5% U
T
(>95% dip in U
for 0.5 cycle
40% U
T
(60% dip in U
for 5 cycles
70% U
T
(30% dip in U
for 25 cycles
<5% U
T
(>95% dip in U
for 5 sec
3 A/m
is the a.c. mains voltage prior to application of the test level.
T
C
EsT
omplianCE
l
EvEl
±(2,4,6) kV
contact
±(2,4,8) kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV
differential
mode
±2 kV for
common mode
<5% U
T
)
(>95% dip in
T
U
) for 0.5 cycle
T
40% U
T
)
(60% dip in U
)
T
T
for 5 cycles
70% U
T
)
(30% dip in U
)
T
T
for 25 cycles
<5% U
T
)
(>95% dip in
T
U
) for 5 sec
T
3 A/m
E
lECTromagnETiC
E
- g
nvironmEnT
uidanCE
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should
be that of a typical commercial
or hospital environment. If the
user of the goLITE BLU requires
continued operation during
power mains interruptions,
it is recommended that the
goLITE BLU be powered from an
uninterruptible power supply or
a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.