Guidance and manufacturer's declaration – electromagnetic immunity
The LMD-X2705MD/X2700MD is intended for use in the electromagnetic environment specified below. The customer
or the user of the LMD-X2705MD/X2700MD should assure that it is used in such an environment.
IEC 60601
Immunity test
test level
Electrostatic
±8 kV contact
discharge (ESD)
±15 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV line(s) to
line(s)
IEC 61000-4-5
±2 kV line(s) to
earth
Voltage dips,
0%
T
short
(100% dip in
interruptions
)
T
and voltage
for 0.5/1 cycles
variations on
power supply
40%
T
input lines
(60% dip in
for 5 cycles
IEC 61000-4-11
70%
T
(30% dip in
for 25/30
a
cycles
(for 0.5 sec)
0%
T
(100% dip in
)
T
for 250/300
a
cycles
(for 5 sec)
Power
30 A/m
frequency (50/
60 Hz) magnetic
field
IEC 61000-4-8
NOTE:
is the a.c. mains voltage prior to application of the test level.
T
a For example, 10/12 means 10 cycles at 50 Hz or 12 cycles at 60 Hz.
Compliance level
AC input
DC input
±8 kV contact
±8 kV contact
±15 kV air
±15 kV air
±2 kV for power
supply lines
±1 kV for input/
±1 kV for input/
output lines
output lines
±1 kV
Not applicable
differential
mode
±2 kV common
mode
0%
Not applicable
T
(100% dip in
)
T
a
for 0.5/1 cycles
a
40%
T
)
(60% dip in
)
T
T
for 5 cycles
70%
T
)
(30% dip in
)
T
T
for 25/30
a
cycles
(for 0.5 sec)
0%
T
(100% dip in
)
T
for 250/300
a
cycles
(for 5 sec)
30 A/m
30 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, a relative humidity of at least 30% is
recommended.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of the LMD-X2705MD/X2700MD
requires continued operation during power
mains interruptions, it is recommended that
the LMD-X2705MD/X2700MD be powered
from an uninterruptible power supply or a
battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.